【数据集背景】
本数据集源站标题为《Ray propagation imaging and optical quality evaluation of different intraocular lens models》。配套论文《Ray propagation imaging and optical quality evaluation of different intraocular lens models》发表于 PLOS ONE(2020),DOI:10.1371/journal.pone.0228342。源站首次发布:2020-02-04。
【数据内容】
源站原始描述:Purpose Ray propagation visualization and optical performance analysis of four different intraocular lenses (IOLs) Methods In this laboratory study, four IOLs with different optical designs were assessed: a monofocal AcrySof IQ SN60WF [Alcon], a diffractive-refractive bifocal AcrySof IQ Restor SN6AD1 [Alcon], a diffractive trifocal AcrySof IQ PanOptix TFNT00 [Alcon], and a diffractive extended-depth-of-focus (EDOF) Symfony ZXR00 [Johnson&Johnson]. An experimental set-up with a water bath contain…
【数据结构与技术规格】
文件数 2 个,合计 180.34 KB。pone.0228342.s001.xlsx:application/vnd.openxmlformats-officedocument.spreadsheetml.sheet,165.74 KB。pone.0228342.s002.docx:application/vnd.openxmlformats-officedocument.wordprocessingml.document,14.60 KB。数据集 DOI:10.1371/journal.pone.0228342。
【主题与分类】
源站学科分类:Physical Sciences not elsewhere classified、Cell Biology、Physiology、Biotechnology、Chemical Sciences not elsewhere classified、Plasma Physics。主题标签:EDOF IOL、Through-Focus Response、modulation transfer function、diffractive extended-depth-of-focus、ray propagation imaging、Trioptics GmbH、monofocal AcrySof IQ SN 60WF、ray propagation visualization technique、ray propagation behavior、Alcon。
【适用方向】
多模态数据适合跨模态对齐、融合建模与联合表征学习方向的研究。
【获取与许可】
源站页面:https://plos.figshare.com/articles/dataset/Ray_propagation_imaging_and_optical_quality_evaluation_of_different_intraocular_lens_models/11801730
使用许可:CC BY 4.0(https://creativecommons.org/licenses/by/4.0/)。
规范引用:Son, Hyeck Soo; Labuz, Grzegorz; Khoramnia, Ramin; Merz, Patrick; Yildirim, Timur M.; Auffarth, Gerd U. (2020). Ray propagation imaging and optical quality evaluation of different intraocular lens models. PLOS ONE. Dataset
【使用注意】
· 源站最后修改于 2023-05-30,引用时请注意版本时效。
下载信息
注册下载
需要注册 Kaggle 账号并登录后下载,适合需要跟踪下载记录和使用 API 的用户。
暂未开放公开下载
无需注册即可直接获取公开样本或文档,适合快速预览和评估数据集质量。
免登录有偿下载
公开数据集受托下载与技术交付服务。
提供高速下载与技术交付服务(收技术服务费,非数据销售)
暂未开放当前数据集主要来源为 Kaggle 公开托管,完整影像包建议通过原始链接或 Kaggle API 获取。
使用方式
数据集获取(Figshare)
- 源站标题:Ray propagation imaging and optical quality evaluation of different intraocular lens models
- DOI:10.1371/journal.pone.0228342
- 发布日期:2020-02-04
- 文件数:2 个
- 文件总体积:180.34 KB
- 源站页面:https://plos.figshare.com/articles/dataset/Ray_propagation_imaging_and_optical_quality_evaluation_of_different_intraocular_lens_models/11801730
命令行下载
curl -L -o "pone.0228342.s001.xlsx" "https://ndownloader.figshare.com/files/21539775"
curl -L -o "pone.0228342.s002.docx" "https://ndownloader.figshare.com/files/21539778"
文件清单
pone.0228342.s001.xlsx(165.74 KB)pone.0228342.s002.docx(14.60 KB)
文件总体积 180.34 KB(源站 API 实测)。
数据说明
Purpose
Ray propagation visualization and optical performance analysis of four different intraocular lenses (IOLs)
Methods
In this laboratory study, four IOLs with different optical designs were assessed: a monofocal AcrySof IQ SN60WF [Alcon], a diffractive-refractive bifocal AcrySof IQ Restor SN6AD1 [Alcon], a diffractive trifocal AcrySof IQ PanOptix TFNT00 [Alcon], and a diffractive extended-depth-of-focus (EDOF) Symfony ZXR00 [Johnson&Johnson]. An experimental set-up with a water bath containing 0.01% fluorescein solution and monochromatic green laser light (532 nm) was used to visualize the propagation of light rays. Also, the optical performance of the IOLs was evaluated by measuring the modulation transfer function (MTF) values at a pupil sizes of 3.0 and 4.5 mm on the optical bench OptiSpheric® IOL PRO II (Trioptics GmbH, Germany).
Results
Both the diffractive-refractive bifocal IOL and the EDOF IOL showed two defined foci for distance and near vision. In the diffractive trifocal IOL, three distinct foci for distance, intermediate, and near vision could be visualized.
Conclusions
The ray propagation visualization technique allows a qualitative assessment and comparison of light energy distribution between different IOL models. The measured Through-Focus Response (TFR) quantitatively confirmed the evaluated ray propagation behavior.