基于ESHO-QA指南,通过红外相机测量体模温度分布,评估LCA浅表热疗加热系统性能的公开数据集。

基本信息
资源简介
应用ESHO-QA指南评估LCA浅表热疗加热系统性能
【一句话概览】
本数据集是 figshare 公开的影像数据集,收录 1 个文件、合计 2.80 MB,主题涵盖 Hyperthermia、quality assurance (QA)、superficial applicators、LCA,可用于医学图像分割、检测与分类模型的训练与评测,CC BY 4.0 许可开放获取。
【数据集背景】
本数据集源站标题为《Application of the ESHO-QA guidelines for determining the performance of the LCA superficial hyperthermia heating system》。配套论文《Application of the ESHO-QA guidelines for determining the performance of the LCA superficial hyperthermia heating system》发表于 International Journal of Hyperthermia(2023),DOI:10.1080/02656736.2023.2272578。项目资助方:This research has received support from the European Union’s Horizon 2020 research and innovation program under the Marie Skłodowska-Curie (MSCA-ITN) grant ‘Hyperboost’ project, no. 955625.。源站首次发布:2023-10-26。
【数据内容】
源站原始描述:This study aimed to assess the quality of the lucite cone applicator (LCA), the standard applicator for superficial hyperthermia at the Erasmus MC Cancer Institute, using the most recent quality assurance guidelines, thus verifying their feasibility. The assessment was conducted on each of the six LCAs available for clinical treatments. The temperature distribution was evaluated using an infrared camera across different layers of a fat-muscle mimicking phantom. The maximum temperature increase,…
【数据结构与技术规格】
文件数 1 个,合计 2.80 MB。ihyt_a_2272578_sm5951.xlsx:application/vnd.openxmlformats-officedocument.spreadsheetml.sheet,2.80 MB。数据集 DOI:10.6084/m9.figshare.24438740.v1。
【主题与分类】
源站学科分类:Space Science、Medicine、Cell Biology、Biotechnology、Chemical Sciences not elsewhere classified、Biological Sciences not elsewhere classified。主题标签:Hyperthermia、quality assurance (QA)、superficial applicators、LCA。
【适用方向】
影像类数据可用于医学图像分析模型(检测、分割、分类)的训练与评测,也可用于标注质量与成像参数相关性的研究。
【获取与许可】
源站页面:https://tandf.figshare.com/articles/dataset/Application_of_the_ESHO-QA_guidelines_for_determining_the_performance_of_the_LCA_superficial_hyperthermia_heating_system/24438740
使用许可:CC BY 4.0(https://creativecommons.org/licenses/by/4.0/)。
规范引用:Application of the ESHO-QA guidelines for determining the performance of the LCA superficial hyperthermia heating system. Taylor & Francis. Carrapiço-Seabra, Carolina; De Lazzari, Mattia; Ameziane, Abdelali; van Rhoon, Gerard C.; Trefná, Hana Dobšícek; Curto, Sergio
【使用注意】
· 源站最后修改于 2024-03-21,引用时请注意版本时效。
下载信息
注册下载
需要注册 Kaggle 账号并登录后下载,适合需要跟踪下载记录和使用 API 的用户。
暂未开放公开下载
无需注册即可直接获取公开样本或文档,适合快速预览和评估数据集质量。
免登录有偿下载
公开数据集受托下载与技术交付服务。
提供高速下载与技术交付服务(收技术服务费,非数据销售)
暂未开放当前数据集主要来源为 Kaggle 公开托管,完整影像包建议通过原始链接或 Kaggle API 获取。
使用方式
数据集获取(Figshare)
- 源站标题:Application of the ESHO-QA guidelines for determining the performance of the LCA superficial hyperthermia heating system
- DOI:10.6084/m9.figshare.24438740.v1
- 发布日期:2023-10-26
- 文件数:1 个
- 文件总体积:2.80 MB
- 源站页面:https://tandf.figshare.com/articles/dataset/Application_of_the_ESHO-QA_guidelines_for_determining_the_performance_of_the_LCA_superficial_hyperthermia_heating_system/24438740
命令行下载
curl -L -o "ihyt_a_2272578_sm5951.xlsx" "https://ndownloader.figshare.com/files/42902489"
文件清单
ihyt_a_2272578_sm5951.xlsx(2.80 MB)
文件总体积 2.80 MB(源站 API 实测)。
数据说明
This study aimed to assess the quality of the lucite cone applicator (LCA), the standard applicator for superficial hyperthermia at the Erasmus MC Cancer Institute, using the most recent quality assurance guidelines, thus verifying their feasibility.
The assessment was conducted on each of the six LCAs available for clinical treatments. The temperature distribution was evaluated using an infrared camera across different layers of a fat-muscle mimicking phantom. The maximum temperature increase, thermal effective penetration depth (TEPD), and thermal effective field size (TEFS) were used as quality metrics. The experimental results were validated through comparison with simulated results, using a canonical phantom model and a realistic phantom model segmented from CT imaging.
A maximum temperature increase above 6 °C at 2 cm depth in the fat-muscle phantom for all the experiments was found. A mean negative difference between simulated and experimental data was of 1.3 °C when using the canonical phantom model. This value decreased to a mean negative difference of 0.4 °C when using the realistic model. Simulated and measured TEPD showed good agreement for both in silico scenarios, while discrepancies were present for TEFS.
The LCAs passed all QA guidelines requirements for superficial hyperthermia delivery when used singularly or in an array configuration. A further characterization of parameters such as antenna efficiency and heat transfer coefficients would be beneficial for translating experimental results to simulated values. Implementing the QA guidelines was time-consuming and demanding, requiring careful preparation and correct setup of antenna elements.